Difference between revisions of "Projects/Diffusion/2007 Project Week Contrasting Tractography Measures/Acquisition Parameters"
From NAMIC Wiki
RandyGollub (talk | contribs) |
|||
(2 intermediate revisions by 2 users not shown) | |||
Line 1: | Line 1: | ||
DTI data: | DTI data: | ||
+ | |||
+ | 3 Tesla Dataset | ||
* 3 Tesla GE system was used to acquire data (General Electric Medical Systems, Milwaukee, WI) | * 3 Tesla GE system was used to acquire data (General Electric Medical Systems, Milwaukee, WI) | ||
* We used Echo Planar Imaging (EPI) DTI Tensor acquisiton with double echo option enabled, in order to reduce eddy-current related distortions. | * We used Echo Planar Imaging (EPI) DTI Tensor acquisiton with double echo option enabled, in order to reduce eddy-current related distortions. | ||
* 8 Channel Head Coil was used to reduce the impact of EPI spatial distortions, and ASSET (Array Spatial Sensitivity Encoding Techniques, GE) with a SENSE-factor (speed-up) of 2 was used. | * 8 Channel Head Coil was used to reduce the impact of EPI spatial distortions, and ASSET (Array Spatial Sensitivity Encoding Techniques, GE) with a SENSE-factor (speed-up) of 2 was used. | ||
* Acquisition parameters were as follows: 51 directions with b=700, 8 baseline scans with b=0, TR 17000 ms, TE 78 ms, FOV 24 cm, 144 x 144 encoding steps, 1.7 mm slice thickness. The total scan time for the sequence was 17 minutes. | * Acquisition parameters were as follows: 51 directions with b=700, 8 baseline scans with b=0, TR 17000 ms, TE 78 ms, FOV 24 cm, 144 x 144 encoding steps, 1.7 mm slice thickness. The total scan time for the sequence was 17 minutes. | ||
+ | |||
+ | MIND Dataset | ||
+ | * 1.5 T Siemens using 8-Channel Coil | ||
+ | * 60 gradient directions, 10 baselines, b = 700. | ||
+ | * TR = 8900 ms, TE = 80 ms, FOV = 256 mm, 2.0 mm voxel size, 2 mm slice thickness, 128 x 128 x 64 matrix size. | ||
+ | * Scan time: 10:32 min | ||
+ | |||
− | Return to [[Projects/Diffusion/2007_Project_Week_Contrasting_Tractography_Measures | | + | Return to [[Projects/Diffusion/2007_Project_Week_Contrasting_Tractography_Measures | Contrasting Tractography Project Page]] |
Latest revision as of 20:46, 24 January 2008
Home < Projects < Diffusion < 2007 Project Week Contrasting Tractography Measures < Acquisition ParametersDTI data:
3 Tesla Dataset
- 3 Tesla GE system was used to acquire data (General Electric Medical Systems, Milwaukee, WI)
- We used Echo Planar Imaging (EPI) DTI Tensor acquisiton with double echo option enabled, in order to reduce eddy-current related distortions.
- 8 Channel Head Coil was used to reduce the impact of EPI spatial distortions, and ASSET (Array Spatial Sensitivity Encoding Techniques, GE) with a SENSE-factor (speed-up) of 2 was used.
- Acquisition parameters were as follows: 51 directions with b=700, 8 baseline scans with b=0, TR 17000 ms, TE 78 ms, FOV 24 cm, 144 x 144 encoding steps, 1.7 mm slice thickness. The total scan time for the sequence was 17 minutes.
MIND Dataset
- 1.5 T Siemens using 8-Channel Coil
- 60 gradient directions, 10 baselines, b = 700.
- TR = 8900 ms, TE = 80 ms, FOV = 256 mm, 2.0 mm voxel size, 2 mm slice thickness, 128 x 128 x 64 matrix size.
- Scan time: 10:32 min