Difference between revisions of "2008 IGT Project Week Prostate Robotics"
From NAMIC Wiki
Line 61: | Line 61: | ||
*Results of the experiments | *Results of the experiments | ||
− | + | Software functions | |
{| border="1" cellpadding="5" cellspacing="0" align="center" | {| border="1" cellpadding="5" cellspacing="0" align="center" | ||
Line 92: | Line 92: | ||
<font color="red">* ... implemented during the IGT Week </font> | <font color="red">* ... implemented during the IGT Week </font> | ||
− | + | Experiments | |
**We followed clinical procedures | **We followed clinical procedures | ||
**We performed needle insertion into prostate phantom | **We performed needle insertion into prostate phantom |
Revision as of 05:34, 12 December 2008
Home < 2008 IGT Project Week Prostate RoboticsBack to 2008_IGT_Project_Week#Projects
Team: Noby Hata, Clif Burdette, Jack Blevins, Clare Tempany, Sam Song, Bong Joon Cho, Greg Fisher, Roland Goerlitz, Junichi Tokuda
Contents
Goals
- Mock-up biopsy and therapy session
- Discuss robot control through OpenIGTLink
Place
Approach
Schedule
8 Monday | AM | AMS and JHU teams travel to Boston |
PM | Workshop 3:30pm-4:30pm Prostate breakout session = "planning for the week" | |
7:20pm - 12:00am | MRI @ 221 Longwood: Experiments. Goal all interface are up running. Checking communication with MRI and interdependent communication. Robot registration. | |
9 Tuesday | AM | Workshop: Junici-Jack realtime image transfer |
PM | Workshop: Jack-JHU definition of the IGT formats. | |
10 Wednesday | AM | Workshop : experiment design? |
PM | Workshop | |
7:20pm - 12:00am | MRI @ 221 Longwood: Experiments "slow mock-up MRg biopsy and accuracy assessment -- Phantom -- rehearsal for Thursday" | |
11 Thursday | AM | Workshop: Discussion for 2nd generation Robot. |
PM | Workshop | |
7:20pm - 12:00am | MRI @ 221 Longwood: Experiments "mock-up Slicer-based biopsy AND RadVision-based brachy" | |
12 Friday | AM | Workshop |
PM | -- |
http://wiki.ncigt.org/index.php/U41:Prostate:BRP:MRI_Robot_Experiment_2008_December
Results
- Results of the experiments
Software functions
Function | Slicer | RadVision |
Workphase command | OK | OK |
"SET_Z_FRAME" command | OK* | OK |
"MOVE_TO" command | OK* | OK |
"GET_COORDINATE" command | OK* | OK |
"GET_STATUS" command | not implemented | OK |
"HOME" command | not implemented | not tested |
"INSERT" command | not implemented | not implemented |
"BIOPSY" command | not implemented | not implemented |
Real-time image transfer | OK | OK* |
Real-time image control | OK | OK*(orientation is not supported) |
Z-frame calibration | implemented* (with visualization) | implemented (not integrated with comunication) |
* ... implemented during the IGT Week
Experiments
- We followed clinical procedures
- We performed needle insertion into prostate phantom
- We confirmed that the needle is inserted around the targeted position on the real-time image
- Robot actuator was not used due to minor mechanical trouble. (robot was moved manually)
- Pulse sequence and parameters should be carefully chosen for Z-frame calibration
- Plan for the next