Difference between revisions of "MBIRN: structural calibration dissemination"

From NAMIC Wiki
Jump to: navigation, search
m (Update from Wiki)
 
m (Update from Wiki)
 
(2 intermediate revisions by the same user not shown)
Line 7: Line 7:
 
<br />
 
<br />
  
* '''Human Acquisition Protocol'''
+
* [[Media:MBIRN-scan-summary-table.pdf| mBIRN data dissemination overview]]
 
 
Sequence: 3D spoiled gradient echo (FLASH type)
 
Vendor's standard birdcage RF coil
 
Slice orientation: Sagittal
 
No partial Fourier
 
NEX=1
 
FOV= 25cm
 
Phase matrix=192
 
Readout matrix=256
 
Number of Slices: 124 (no skip, whole head coverage, avoid wrap around)
 
Slice Thickness: 1.3mm-1.5mm
 
TE= minimum allowed with centered (symmetric) echo.
 
TR=20 ms
 
Slab Selection: No (i.e., hard RF pulses)
 
scan 1: flip angle 30 degrees  (T1-weighted image)
 
scan 2: flip angle 5 degrees (proton-density-weighted image)
 
scan 3: flip angle 20 degrees (T1-weighted image)
 
scan 4: flip angle 3 degrees  (proton-density-weighted image)
 
Acquisition time per scan: about 8min 12sec
 
 
 
<br />
 
 
 
* '''Phantom Acquisition Protocol'''
 
 
 
Sequence: 3D spoiled gradient echo (FLASH type)
 
Vendor's standard birdcage RF coil
 
Slice orientation: Sagittal
 
No partial Fourier
 
NEX=1
 
FOV= 40cm
 
Phase matrix=192
 
Readout matrix=256
 
Number of Slices: 124 (no skip, whole phantom coverage, avoid wrap around)
 
Slice Thickness: 2 mm
 
TE= minimum allowed with centered (symmetric) echo.
 
TR=20 ms
 
Slab Selection: No (i.e., hard RF pulses)
 
scan 1: flip angle 30 degrees without vendor's distortion correction
 
scan 2: flip angle 30 degrees with vendor's distortion correction (if available)
 
Acquisition time per scan: about 8min 12sec
 
  
 
<br />
 
<br />

Latest revision as of 14:04, 18 December 2006

Home < MBIRN: structural calibration dissemination
  • Overview of dissemination contents
    • Multi-site phantom data: phantom that helps asses image distortions from gradient non-linearities
    • Multi-site human data: healthy volunteers, most of which were scanned twice at the participating sites
    • Human Protocol: multi-spectral 3D structural (FLASH with multiple flip angles: 30, 20, 5, 3 degrees)
    • Sites: Five sites with different 1.5T systems (Siemens, GE, Philips/Picker)




Back to: Mbirn: Structural MRI calibration data dissemination