Difference between revisions of "2013 Winter Project Week Gyne Brachy"

From NAMIC Wiki
Jump to: navigation, search
m (Text replacement - "http://www.slicer.org/slicerWiki/index.php/" to "https://www.slicer.org/wiki/")
 
(5 intermediate revisions by one other user not shown)
Line 3: Line 3:
 
Image:PW-SLC2013.png|[[2013_Winter_Project_Week#Projects|Projects List]]
 
Image:PW-SLC2013.png|[[2013_Winter_Project_Week#Projects|Projects List]]
 
Image:needles.jpg| Segmented Brachytherapy Needles in iGyne.
 
Image:needles.jpg| Segmented Brachytherapy Needles in iGyne.
 +
Image:01.png| MRI to CT registration.
 +
Image:011.png| Segmented brachytherapy Needles in MRI with patient lithotomy position (green) versus segmented needles from CT Scnas in supine position (red).
 
</gallery>
 
</gallery>
  
Line 15: Line 17:
 
<div style="width: 27%; float: left; padding-right: 3%;">
 
<div style="width: 27%; float: left; padding-right: 3%;">
 
<h3>Objective</h3>
 
<h3>Objective</h3>
* The [http://www.slicer.org/slicerWiki/index.php/Documentation/Nightly/Extensions/iGyne iGyne module] is a user friendly Slicer module which is designed to assist gynecologic radiation oncologists in the process of applicator placement and brachytherapy needle insertion. Currently the module can perform applicator registration and needle segmentation successfully.
+
* The [https://www.slicer.org/wiki/Documentation/Nightly/Extensions/iGyne iGyne module] is a user friendly Slicer module which is designed to assist gynecologic radiation oncologists in the process of applicator placement and brachytherapy needle insertion. Currently the module can perform applicator registration and needle segmentation successfully.
 
* The objective of this project is to develop a method to validate and measure the accuracy of the needle segmentation results.  
 
* The objective of this project is to develop a method to validate and measure the accuracy of the needle segmentation results.  
 
</div>
 
</div>
Line 27: Line 29:
 
<div style="width: 27%; float: left; padding-right: 3%;">
 
<div style="width: 27%; float: left; padding-right: 3%;">
 
<h3>Progress</h3>
 
<h3>Progress</h3>
* To compare the segmented needles from MRI and CT, we tried different landmarks to for registration and we conclude that the best rigid anatomical landmark for this purpose is the hip bone.
+
* To compare the segmented needles from MRI and CT, we tried different landmarks for registration and finally we conclude that the best rigid anatomical landmark for our purpose is the hip bone.
* We found out that the needles displaced noticeably and their shapes has been changed by the patients position change from partial lithotomy position (the MRI data) to supine position (the CT data). This implies that in order to be able to use the CT data for validation iGyne  needle segmentation tool, we need to modify our workflow and add a supine position MRI to our IGT Gyne workflow.
+
* We found out that the needles displaced noticeably and their shapes has been changed by the patients position switch from partial lithotomy position (the MRI data) to supine position (the CT data). This implies that in order to use the CT data for validating the iGyne  needle segmentation tool, we need to modify our workflow and add a supine position MRI to our IGT gynecological brachytherapy workflow.
 
* Our findings about the needle position and shape changes demands a further investigation to understand and model those changes and their possible effect on the treatment results.
 
* Our findings about the needle position and shape changes demands a further investigation to understand and model those changes and their possible effect on the treatment results.
 
* For the purpose of needle segmentation results validation we will add 3D point correspondence and similarity measures to iGyne module soon.
 
* For the purpose of needle segmentation results validation we will add 3D point correspondence and similarity measures to iGyne module soon.
 
</div>
 
</div>
 
</div>
 
</div>

Latest revision as of 18:07, 10 July 2017

Home < 2013 Winter Project Week Gyne Brachy

Key Investigators

  • Alireza Mehrtash, Tina Kapur, Guillaume Pernelle, Jan Egger, Neha Agrawal, Antonio Damato, BWH
  • Csaba Pinter, Queen's

Project Description

Objective

  • The iGyne module is a user friendly Slicer module which is designed to assist gynecologic radiation oncologists in the process of applicator placement and brachytherapy needle insertion. Currently the module can perform applicator registration and needle segmentation successfully.
  • The objective of this project is to develop a method to validate and measure the accuracy of the needle segmentation results.

Approach, Plan

  • Needle digitization.
  • Exporting the segmented needles.
  • Make a comparison between the segmented needles from MRI images and the postoperative checking CT images.
  • Define metrics to measure the accuracy of the segmentation results.

Progress

  • To compare the segmented needles from MRI and CT, we tried different landmarks for registration and finally we conclude that the best rigid anatomical landmark for our purpose is the hip bone.
  • We found out that the needles displaced noticeably and their shapes has been changed by the patients position switch from partial lithotomy position (the MRI data) to supine position (the CT data). This implies that in order to use the CT data for validating the iGyne needle segmentation tool, we need to modify our workflow and add a supine position MRI to our IGT gynecological brachytherapy workflow.
  • Our findings about the needle position and shape changes demands a further investigation to understand and model those changes and their possible effect on the treatment results.
  • For the purpose of needle segmentation results validation we will add 3D point correspondence and similarity measures to iGyne module soon.