Projects:DicomToNrrdForDTI

From NAMIC Wiki
Revision as of 16:56, 29 April 2008 by Sgouttard (talk | contribs)
Jump to: navigation, search
Home < Projects:DicomToNrrdForDTI

Dicom conversion for DTI data


Introduction

In this page we present some results on the conversion from dicom data to DTI. Working with this new dataset, we noticed differences between the dicom header information and the ones from the original protocol designed for this study. In this page, we want describe and discuss these differences.

The data we worked with have been acquired with a Siemens 3T Tim Trio.

The DTI sequence is a 25 directions with variable b-values plus one B0 image.

OrigDataBVal.png