Inter-slice Motion Correction for fMRI

From NAMIC Wiki
Jump to: navigation, search
Home < Inter-slice Motion Correction for fMRI

Objective

   * To perform inter-slice motion correction in fMR images using affine and non-rigid registration methods. 


Progress

   * Implemented and tested b-Spline/affine registration with Mattes-MI and KL metrics. 
   * Coding completed. However, results are far from satisfactory.
   * Progress (maybe) stymied by "poor" resolution of data.
   * Progress (maybe) stymied by "not-so-viable" metrics (MI, etc.).

Algorithms Used From ITK

Metrics

  1. Mean Squared - smooth, and for rigid registration, the minimum is fairly correct (physically)
  2. MI (Wells, Mattes) - very non-smooth - lot of discontinuities. For all motions, results are wrong (possibly because cannot identify a good minimum pt)
  3. KL-Divergence (Chung, Wells, et. al.) - again very non-smooth. Cannot identify a good minimum point.

Transformations

  1. Rigid - works well for ms metric, not so correct for statistical metrics.
  2. Affine - works for ms metric, does not work for the statistical metrics.
  3. B-spline non-rigid - tested with ms and kl. Extremely poor results.

Results

Original fMRI

Unreg.png

MS Metric with rigid transform

Ms rigid.png


MS Metric with affine transform

Ms aff.png

Mattes MI metric with rigid transform

Mattes ridid.png

KL metric with rigid transform

Kl rigid.png

SPM Registered

Spm reg.png


Issues

  • As can be seen any registration result looks more jagged (along the sagittal and coronal sections).
  • Low sampling rate and image noise - all the metrics are extremely non-smooth as a result.
  • Even the slightest change in initialization results in a large deviation in the registration result.

Open questions

   * Intensity normalization / histogram equalization - would these have any impact on inter-slice registration? Would they affect the computation of the joint-entropies for MI/KL metrics.
   * Should there be some regularization factor on the registration to limit the motion ?

To Do

   * Right now, registration is done on a slice by slice basis. However, due to the low resolution of fMRI and the absence of gradients result in low registration accuracy. We are investigating using alternative image-to-image metrics like the KL divergence. We are also looking at simultaneously registering multiple slices.

Key Investigators

   * Firdaus Janoos, Raghu Machiraju, Steve Pieper, Wendy Plesniak.