2010 Summer Project Week Prostate Intervention

From NAMIC Wiki
Revision as of 14:18, 25 June 2010 by Tokuda (talk | contribs)
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to: navigation, search
Home < 2010 Summer Project Week Prostate Intervention


Key Investigators

  • Queen's University: Tamas Ungi
  • Haiying Liu: BWH
  • Junichi Tokuda: BWH
  • Sam S. Song: JHU

Objective

Develop an MR-compatible needle placement manipulator for MR-guided prostate intervention.

Approach, Plan

We will perform a mock needle biopsy and brachytherapy using the new device at Brigham and Women's Hospital. 3D Slicer is used for planning, guidance and robot control. This is a continued work from 2009 Project Week.

Progress

  • We performed Mock therapy in 3T MRI Scanner at Brigham and Women's Hospital
  • Fix problem in sharing z-frame registration data among 3D Slicer and the robot
  • Fix registration error caused by changed z-frame geometry
  • All fix are uploaded to NA-MIC sand box repository (IGTLoadableModules/ProstateNavBWH). This will be merged to main repository (IGTLoadableModules/ProstateNav).

Delivery Mechanism

This work will be delivered to the NA-MIC Kit as a (please select the appropriate options by noting YES against them below)


References