Difference between revisions of "Projects:RegistrationLibrary:RegLib C17"
From NAMIC Wiki
(Created page with 'Back to ARRA main page <br> Back to Registration main page <br> [[Projects:RegistrationDocumentation:UseCaseInv…') |
|||
Line 40: | Line 40: | ||
=== Procedures=== | === Procedures=== | ||
*''Phase 1'': obtain masks for kidney ROI | *''Phase 1'': obtain masks for kidney ROI | ||
− | #Use the ''Draw'' tool in the | + | #Use the ''Draw'' tool in the [http://www.slicer.org/slicerWiki/index.php/Modules:Editor-Documentation-3.6 Editor] and/or use the [http://www.slicer.org/slicerWiki/index.php/Modules:FastMarchingSegmentation-Documentation-3.6 Fast-Marching Segmentation] module to obtain segmentations of the kidney in both CT and MR image. In the example dataset, you can find those segementations as files "Probe1CT_seg.nrrd" and "MRl_seg.nrrd" |
+ | #See the [http://www.slicer.org/slicerWiki/index.php/Slicer_3.6:Training tutorials] for the Editor and and segmentation methods for details on how to obtain an efficient segmentation. | ||
+ | *''Phase 2'': obtain initial manual alignment | ||
+ | **Details in the [http://www.slicer.org/slicerWiki/index.php/Slicer3.6:Training manual registration tutorial] and also in the [http://www.slicer.org/slicerWiki/index.php/Slicer-3-6-FAQ#How_do_I_initialize.2Falign_images_with_very_different_orientations_and_no_overlap.3F Slicer FAQ] | ||
+ | #go to the [http://www.slicer.org/slicerWiki/index.php/Modules:Data-Documentation-3.6 Data module] | ||
+ | #right click on the "Scene" mode and select ''Insert New Transform''. In the MRML edit field below, rename it to "Xf0_manual" or similar | ||
+ | #In the ''Data'' module, drag the "MRI" image volume inside the"Xf0_manual" node | ||
+ | #Select the views so that the MR and CT volumes are displayed in the slice views | ||
+ | #Go to the [http://www.slicer.org/slicerWiki/index.php/Modules:Transforms-Documentation-3.6 Transforms module] and adjust the translation and rotation sliders to adjust the current position. To get a finer degree of control, enter smaller numbers for the translation limits and enter rotation angles numerically in increments of a few degrees at a time | ||
+ | *''Phase 3'': automated rigid/affine alignment | ||
+ | #Go to the [http://www.slicer.org/slicerWiki/index.php/Modules:BRAINSFit '''BRAINSfit'''] module | ||
+ | ##select Presets "Xf1_Rigid" or "Xf2_Affine" or set the parameters as given below: | ||
+ | ##fixed image: "Probe1CT", moving image: "MRI" | ||
+ | ##''Initialize with previous transform'': select "Xf0_manual" from Phase 1 above | ||
+ | ##''Initialize Transform Mode'': leave at default = OFF | ||
+ | ##check ''Include Rigid registration Phase'' box. For affine also check ''ScaleVersor3D'' and ''Affine'' | ||
+ | ##Output: under ''Slicer Linear Transform'', select new and rename to "Xf1_Rigid" or "Xf2_Affine" or similar | ||
+ | ##Registration Parameters: set "Number of Samples" to 100,000 at least | ||
+ | ##Mask Processing: check ''ROI'' box, ''Input Fixed Mask'' select CT segmentation created in Phase 1 above. For ''Input Moving Mask'' select MRI segmentation created in Phase 1 above. | ||
+ | ##Click Apply | ||
+ | #return to the [http://www.slicer.org/slicerWiki/index.php/Modules:Data-Documentation-3.6 Data module] and drag the MRI inside/outside the different registration transforms to compare the alignment | ||
+ | #to obtain a resampled volume: move MRI inside Xform of choice and then right-click on the volume and select ''Harden Transforms''. Save MRI under new name. | ||
+ | |||
=== Registration Results=== | === Registration Results=== | ||
[[Image:RegLib_C17_unregistered.gif||400px|original unregistered]] | [[Image:RegLib_C17_unregistered.gif||400px|original unregistered]] |
Revision as of 15:44, 5 September 2011
Home < Projects:RegistrationLibrary:RegLib C17Back to ARRA main page
Back to Registration main page
Back to Registration Use-case Inventory
Contents
v3.6.3 Slicer Registration Library Case #17:
Kidney pre-op MR to intra-op CT
Input
fixed image/target: intra-operative CT | moving image: per-operative MRI |
Modules
- BRAINSfit
- for mask generation: Draw tool in the Editor and/or Fast-Marching Segmentation,
Keywords
CT, MRI, abdominal, image-guided therapy, IGT, tumor ablation
Input Data
- fixed: intra-operative CT, oblique, 0.58 x 0.58 x 3 mm voxel size, dimensions 512 x 512 x 20
- moving: opre-operative MRI lique, 1.2 x 1.2 x 4 mm voxel size, dimensions 320 x 240 x 64
Registration Challenges
- intra-operative CT is acquired with limited FOV and very oblique position
- intra-operative CT contains artifacts from surgical needles
- both datasets have strong voxel anisotropy
- the region of interest (kidney) is small compared to the FOV, i.e. there is substantial "distracting" image content
Key Strategies
- manual alignment to obtain initial pose
- rough segmentation of both kidneys to mask "distracting" image content
- gradual increase in registration DOF.
Procedures
- Phase 1: obtain masks for kidney ROI
- Use the Draw tool in the Editor and/or use the Fast-Marching Segmentation module to obtain segmentations of the kidney in both CT and MR image. In the example dataset, you can find those segementations as files "Probe1CT_seg.nrrd" and "MRl_seg.nrrd"
- See the tutorials for the Editor and and segmentation methods for details on how to obtain an efficient segmentation.
- Phase 2: obtain initial manual alignment
- Details in the manual registration tutorial and also in the Slicer FAQ
- go to the Data module
- right click on the "Scene" mode and select Insert New Transform. In the MRML edit field below, rename it to "Xf0_manual" or similar
- In the Data module, drag the "MRI" image volume inside the"Xf0_manual" node
- Select the views so that the MR and CT volumes are displayed in the slice views
- Go to the Transforms module and adjust the translation and rotation sliders to adjust the current position. To get a finer degree of control, enter smaller numbers for the translation limits and enter rotation angles numerically in increments of a few degrees at a time
- Phase 3: automated rigid/affine alignment
- Go to the BRAINSfit module
- select Presets "Xf1_Rigid" or "Xf2_Affine" or set the parameters as given below:
- fixed image: "Probe1CT", moving image: "MRI"
- Initialize with previous transform: select "Xf0_manual" from Phase 1 above
- Initialize Transform Mode: leave at default = OFF
- check Include Rigid registration Phase box. For affine also check ScaleVersor3D and Affine
- Output: under Slicer Linear Transform, select new and rename to "Xf1_Rigid" or "Xf2_Affine" or similar
- Registration Parameters: set "Number of Samples" to 100,000 at least
- Mask Processing: check ROI box, Input Fixed Mask select CT segmentation created in Phase 1 above. For Input Moving Mask select MRI segmentation created in Phase 1 above.
- Click Apply
- return to the Data module and drag the MRI inside/outside the different registration transforms to compare the alignment
- to obtain a resampled volume: move MRI inside Xform of choice and then right-click on the volume and select Harden Transforms. Save MRI under new name.